Manual

34
Operation Method
Program Creation Example
11
Press the PROGRAM key to display the "Rep. start".
Rep. start No
S01
12
Press the ENTER key to display the repeat initiating step edition
screen. The cursor goes out and the setting for repeat initiating
step blinks.
Rep. start No
S01
13
Set the "No" using the "▼▲".
Rep. start No
S01
14
Press the ENTER key to decide the setting.
Rep. start No
S01
15
Press the PROGRAM key until the "Set TEMP" in the Step 2 is
displayed. The display on the lower column changes from "S01"
to "S02"or to "SO2 (un-used)".
Set TEMP
0.0
S02
16
Press the ENTER key to display the setting temperature edition
screen. The cursor goes out and the setting temperature blinks.
Set TEMP
0.0
S02
17
Set the temperature to 20 using the "▼▲"
Set TEMP
20.0
S02
18
Press the ENTER key to decide the temperature.
Set TEMP
20.0
S02
19
Press the PROGRAM key to display the "Set time" selection
screen.
Set time 0min
S02
20
Press the ENTER key to display the setting period edition screen.
The cursor goes out and the setting period blinks.
Set time 0min
S02
21
Set the period to 0 min using the "▼▲".
Set time 0min
S02
22
Press the ENTER key to decide the period.
Set time 0min
S02
23
Press the PROGRAM key to display the "Rep. start".
Rep. start No
S02
24
Press the ENTER key to display the repeat initiating step edition
screen. The cursor goes out and the setting for repeat initiating
step blinks.
Rep. start No
S02