User Manual

4
Before Using this unit
Requirements for Installati on
WARNING!
1. Al ways ground this unit
Connect the power plug t o a recept acle w ith grounding connect ors.
Do not forget to ground this unit , t o protect you and t he unit from electrical shock in case of
power surge. Choose a receptacle with grounding connectors as often as possible.
Do not connect the grounding w ire to a gas pipe, or by means of a lightning rod or telephone
line. A fire or elec trical shock will occur.
2. Choose a proper place for installation
Do not install this unit in a place where:
Rough or dirty surface.
Flamm able gas o r corrosive gas is generated.
Ambient temperature exceeds 35°C.
Ambient temperature fluctuates violently.
There is direct sunlight.
There i s excessive humidity and dust.
There is a constant vibration.
Install this unit on a st able place w it h t he space as show n below. Furthermore, make a space
30cm or more above this unit.
Main Unit
More than
15cm
More than
15cm
More than 15cm
Front side
3. Do not use this unit in an area where there is flammable or explosive gas
Never use this unit in an area
where there is flammable or
explosive gas. This unit is not
explosion-proof. An arc may
be generated when the pow er
switch is turned O N or O FF,
and fire/explosion may result.
(To know about flammable or
explosive gas, refer to
page24 “List of Dangerous
Substances”.)