User Manual

21
Specification
Model IJ100
Operating temperature range
0 to 60
(room temp: 20, no load)*
Temperature adjustment accuracy
±0.3 (the center of inner bath, set temp.:37)
Temperature distribution accuracy
±1.0 (the center of inner bath, set temp.:37)
Temperature rising time Approx. 40min. (20 to 60°C)
Performance
Temperature falling time Approx. 100min. (20 to 0°C)
Heating device Mica heater 120W
Cooling device
Two peltier cells, forced radiating method:
It works continuous at the setting temperature below 35
.
(It is turned off at the setting temperature 35
and over.)
Configuration
Internal fan Axial fan (DC)
Controller Hitec IV
FR
type1
Temperature control system PID control of heater output by microcomputer
Temperature setting system
Digital Setting by
▼▲ (up/down) Keys
Operational mode Fixed temperature operation
Sensor Thermistor
Control Section
Additional function
Temperature preset function (1 for desired temperature is able to
set/load)
Self-diagnostic function
Sensor error, Heater error, Triac error, Main relay error, Auto
overheating prevention
Safety
Device
Preservation device
Earth leakage breaker, Independent overheating prevention
device
External dimensions
350(W)
×396(D)×535(H) mm
Internal dimensions
250(W)
×250(D)×250(H) mm
(Air jacket installed: 200
×200×200 mm)
Internal capacity 15.6L
Door
Single door with magnet packing
(the direction of opening/closing is changeable)
Weight Approx. 20kg
Standard
Power supply 100V AC 50/60Hz 3.5A
Accessories
Shelf
×2 (Load: 5kg for each), Plate for drainage, Instruction
manual
Optional accessories
Extra shelf (with metal fittings), Air jack et made of aluminum (with
shelf)
The values shown on “Performance” are in power supply 100V AC.
Usable ambient temperature for this unit is 5 to 30.