User guide

5
2. INPUTTING, EDITION AND DELETING PROGRAMS
2.1 Program Composition
A program is a combination of segments and repeat commands, in which ramp time,
ramp level, repeat frequency, end segment and other parameters should be preset.
[Segment]
Fig 2.1 shows the concept of a segment. Temperature and time patterns of a segment
are determined by the 3 basic parameters below.
Ramp level : Desired temperature
Ramp time : Time required to achieve the preset ramp level
Soak time : Time to hold the ramp level
Fig. 2.1: Concept of A Segment
In Section (
of Fig. 2.1~ temperature varies at a fixed gradient rate toward the
desired temperature set as the ramp level. After reaching the ramp level, temperature
is kept in Section
of Fig 2.1 for the time preset as the soak time. These sections of
and
compose a segment.
If you want to vary the temperature in Section
on a step (for full-power increase or
decrease), instead of a faced gradient rate, set for the ramp time. Fig. 2.2
shows the pattern of a step variation of the ramp time.