2.Operation Method
Preparation
From mounting of the aluminum block to operation start
1. Mounting the aluminum
block
Use the dedicated aluminum block available separately.
The use of an attached dedicated hand grip will prove convenient for
mounting/removal of the aluminum block.
Caution Possible hazard of burn. Use the hand grip when the
aluminum block is sufficiently cooled.
Do not raise the temperature while keeping the hand
grip left attached.
If necessary, a bar thermometer may be inserted into a
thermometer hole (used also for attaching the hand grip) of the
aluminum block.
(Use the thermometer of maximum 200 or more. Since the bar
thermometer is not attached, obtain it separately.)
2. Connecting the power
cord
Connect the power cord to an appropriate receptacle firmly.
The receptacle to be connected must be of a 100V 15A type.
3. Installing the sample To maintain the temperature accuracy, use a sample container
appropriate in size to the sample installation hole of the aluminum
block.
4. Turning ON power
supply
Turn ON the POWER switch of the main body and set the desired
temperature using▲▼ keys of the operation panel.
5. Starting operation Press the START/STOP key of the operation panel for one second,
and the operation begins.
Aluminum block, which is available separately
The following types of aluminum blocks are available for HF100/HF200. Be sure to use the
aluminum block dedicated to the product concerned.
Specify the product code No. when purchasing.
Product code Name/size No. of blocks that
can be set
213173 For 0.5m micro tube 48
213174 For 1.5m micro tube 36
213181 For 12 test tube 36
213182 For 15 test tube 25
213183 For 16.5 test tube 25
213184 For 18 test tube 20
213175 For 21 test tube 12
213176 For 24 test tube 12
213185 For 25 test tube 12
213177 For 30 test tube 8
Can be used
commonly for
HF21/41/61
Hand grip for
aluminum block
Bar thermometer