17
Changing the Program Operation during Fixed Temperature Operation
(Ex. switching to Program 1 during the fixed temperature operation)
Indication after Step/Step Procedure Explanation
1. Push the
MENU
key.
(When starts flashing, push either
key or
key.)
(Ex. Loading Program 1)
Push the
MENU
key twice during the fixed
temperature operation.
The main display flashes either operational program
number or
(auto-stop).
Ex: In the example on the left, (Program 1)
flashes on the display.
If the display flashes
, push either the
key or
the
key to show your desired program number.
The
PROGRAM
lamp starts flashing.
Since the fixed temperature operation is going on,
the
HEAT
lamp keeps flashing, which indicates that
HC200 is under temperature regulation.
2. Push the
ENTER
key.
Push the ENTER key.
The main display flashes
indicating the
operation start time setting mode.
The sub display flashes
3.
Ex., finishing operation after 2 hours and 30
minutes
Push either the
key or the
key.
Push either the
key or the
key and set your desired
time before operation start.
4. Push the
ENTER
key.
Pushing the
ENTER
key changes the operation
mode from the fixed temperature operation to the
program operation.
The main display shows the latest measured
temperature.
The sub display shows the remaining time before
operation start.
The
STANDBY
lamp starts flashing, indicating that
HC200 is ready for operation.
The
PROGRAM
lamp goes from flashing to steady
lit.
5. Starting the program operation.
When the set time is up, the
STANDBY
lamp goes
out and HC200 starts the program operation.
The sub display shows the latest parameters. See
“Changing Sub Display Screens” on page. 22.
(Ex. Preset desired temperature
in use)