17
Changing the Program Operation during Fixed Temperature Operation
(Ex. switching to Program 1 during the fixed temperature operation)
Indication after Step/Step Procedure Explanation
1. Push the
MENU
key.
(When starts flashing, push either
▼
key or
▲
key.)
(Ex. Loading Program 1)
•
Push the
MENU
key twice during the fixed
temperature operation.
⇒
⇒⇒
⇒
The main display flashes either operational program
number or
(auto-stop).
Ex: In the example on the left, (Program 1)
flashes on the display.
•
If the display flashes
, push either the
▼
key or
the
▲
key to show your desired program number.
⇒
⇒⇒
⇒
The
PROGRAM
lamp starts flashing.
⇒
⇒⇒
⇒
Since the fixed temperature operation is going on,
the
HEAT
lamp keeps flashing, which indicates that
HC200 is under temperature regulation.
2. Push the
ENTER
key.
•
Push the ENTER key.
⇒
⇒⇒
⇒
The main display flashes
indicating the
operation start time setting mode.
⇒
⇒⇒
⇒
The sub display flashes
3.
Ex., finishing operation after 2 hours and 30
minutes
Push either the
▼
key or the
▲
key.
•
Push either the
▼
key or the
▲
key and set your desired
time before operation start.
4. Push the
ENTER
key.
⇒
⇒⇒
⇒
Pushing the
ENTER
key changes the operation
mode from the fixed temperature operation to the
program operation.
⇒
⇒⇒
⇒
The main display shows the latest measured
temperature.
⇒
⇒⇒
⇒
The sub display shows the remaining time before
operation start.
⇒
⇒⇒
⇒
The
STANDBY
lamp starts flashing, indicating that
HC200 is ready for operation.
⇒
⇒⇒
⇒
The
PROGRAM
lamp goes from flashing to steady
lit.
5. Starting the program operation.
⇒
⇒⇒
⇒
When the set time is up, the
STANDBY
lamp goes
out and HC200 starts the program operation.
⇒
⇒⇒
⇒
The sub display shows the latest parameters. See
“Changing Sub Display Screens” on page. 22.
(Ex. Preset desired temperature
in use)