Manual
13
Indication after Step/Step Procedure Explanation
4. Push the
ENTER
key.
•
When the program number on the main display has
reached your desired number, push the
ENTER
key.
⇒
⇒⇒
⇒
The controller turns to the start time setting mode,
and the main display flashes
⇒
⇒⇒
⇒
The sub display shows (time).
•
Follow the procedure below and set the operation start
time.
5. Ex., starting operation in 2 hours 30 minutes.
Push either the
▼
key or the
▲
key.
•
Push either the
▼
key or the
▲
key to flash the start
time of the loaded program.
! You can set the start time in the range from 1
second to 999 minutes. However, when the start
time is more than 100 minutes later, seconds cannot
be set.
Ex : 99 hours and 59 minutes later:
100 minutes:
6. Push the
ENTER
key.
•
Push the
ENTER
key.
⇒
⇒⇒
⇒
The main display shows the latest measured
temperature.
⇒
⇒⇒
⇒
The sub display shows the remaining time before
operation starts.
⇒
⇒⇒
⇒
The
STANDBY
lamp starts flashing to indicate that
HC200 is ready for operation.
⇒
⇒⇒
⇒
The
PROGRAM OPERATION
lamp turns from
flashing to steady lit.