Owner's manual

1. Specifications
Time-up Output
Output at the automatic stop and at the end of program
C contact (Relay contact)
Contact capacity
AC250V
3A
    
 :
DC30V
3A
Joint
M4 Screw terminal (Common)
Alarm Output
Output at the occurrence of error
C contact (Relay contact)
Contact capacity
AC250V
3A
     
DC30V
3A
Joint
M4 Screw terminal (Common)
Temperature Output
(Analog)
Voltage output(DC) according to the measured temperature
See table 1 in the next page about output
0
at 0V
Resolution
0.1
Load impedance
100K
Ω
or more
Joint
M4 Screw terminal (Common)
2. Connection
1. Connect firmly the terminal to use (Outputs which was not specified at the order are
unusable)
2. To use the time-up output or the alarm output;
Connect to COM (C) and NO in case the contact is “Closed” at output.
Connect to COM (C) and NC in case the contact is “Opened” at output.
3. when you use the temperature output, employ a shielded line as a connecting cable for
noise prevention, and cut it as short as possible.
Lengthwise Arrangement Crosswise Arrangement