HITECⅣCR OUTPUT Temperature Output Alarm Output Time-up Output Operation Manual Fifth Edition Yamato Scientific America Inc.
1. Specifications Time-up Output Alarm Output Temperature Output (Analog) • Output at the automatic stop and at the end of program • C contact (Relay contact) • Contact capacity:AC250V 3A :DC30V 3A • Joint:M4 Screw terminal (Common) • Output at the occurrence of error • C contact (Relay contact) • Contact capacity:AC250V 3A :DC30V 3A • Joint:M4 Screw terminal (Common) • • • • • Voltage output(DC) according to the measured temperature See table 1 in the next page about output(0℃ at 0V) Resolution:0.
. Caution in use 1. Be sure to turn off the breaker on connection. 2. Use either the time-up and the alarm output under the contact capacity. 3. Be careful not to short-circuit the temperature output terminal. 4. Be sure to connect the temperature output terminal to the recorder etc., with input impedance of 100kΩ or more. 5. Fasten securely the screw attached to terminal for connection.