Manual
40
Operation Method
Program Creation Method
3
Program creation;
Step 3 (S03) to Step 7 (S07)
Set TEMP
► 0℃
S03(un-used)
TROUBLE
ENTER
CANCEL
DOOR
Set TEMP S07
Program OPR
Set TEMP
► 0℃
S07
Set time ► 0min
S07
Fan ► 0
S07
Rep. start ► No
S07
Rep. start S01
S07
Rep.count ► Endless
S07
① Keep pressing the PROGRAM key until "SteP 3" (S03) is
displayed.
② Edit in the same way to Step 6 (S06) and then keep
pressing the PROGRAM key until the Step 7 (S07) is
displayed.
③ Press the ENTER key. The setting temperature
selection screen is displayed.
④ Press the ENTER key. The setting temperature edition
screen is displayed. Set "50℃" using the "▼▲" and
press the ENTER key to decide it.
⑤ Press the PROGRAM key. The setting period selection
screen is displayed.
⑥ Press the ENTER key. The setting period edition
screen is displayed. Set "2h30min" using the "▼▲"
and press the ENTER key to decide it.
⑦ Press the PROGRAM key. The fan rotation speed
selection screen is displayed.
⑧ Press the ENTER key. The fan rotation speed edition
screen is displayed. Set "10" using the "▼▲" and
press the ENTER key to decide it.
⑨ Press the PROGRAM key. The repeat initiating step
selection screen is displayed.
⑩ Press the ENTER key. The repeat initiating step edition
screen is displayed. Set "S01" using the "▼▲" and
press the ENTER key to decide it.
⑪ Press the PROGRAM key. The repeat count selection
screen is displayed.
⑫ Press the ENTER key. The repeat count edition screen
is displayed. Set "1" using the "▼▲" and press the
ENTER key to decide it.