Manual
39
Operation Method
Program Creation Method
1
Program creation; Step 1 (S01)
Fan ► 10
S01
Rep. start ► No
S01
⑥ Press the PROGRAM key. The fan rotation speed
selection screen is displayed.
⑦ Press the ENTER key. The fan rotation speed edition
screen is displayed. Set "10" using the "▼▲" and
press the ENTER key to decide it. (The next selection
screen can be displayed by pressing the ENTER key
when the setting value is indicated in the selection
screen.)
⑧ Press the PROGRAM key. The repeat initiating step
selection screen is displayed.
⑨ Press the ENTER key. The repeat initiating step edition
screen is displayed. Set "No" using the "▼▲" and
press the ENTER key to decide it.
2
Program creation; Step 2 (S02)
TROUBLE
ENTER
CANCEL
DOOR
Set TEMP S02
S02(un-used)
Set TEMP
► 0℃
S02
Set time ► 0min
S02
Fan ► 10
S02
① Keep pressing the PROGRAM key until "SteP2" (S02) is
displayed.
Set TEMP
► 0℃
S02(un-used)
(Displayed only when nothing is input in Step 2)
② Press the ENTER key. The setting temperature
selection screen is displayed.
③ Press the ENTER key. The setting temperature edition
screen is displayed. Set "100℃" using the "▼▲" and
press the ENTER key to decide it.
④ Press the PROGRAM key. The setting period selection
screen is displayed.
⑤ Press the ENTER key. The setting period edition
screen is displayed. Set "30min" using the "▼▲" and
press the ENTER key to decide it.
⑥ Press the PROGRAM key. The fan rotation speed
selection screen is displayed.
⑦ Press the ENTER key. The fan rotation speed edition
screen is displayed. Set "5" using the "▼▲" and press
the ENTER key to decide it.