Manual

30
Operation Method
Input Program; Set the Repeat Initiating Step
1
Repeat initiating step setting procedure
Select the "Rep. start" and press the ENTER key. The setting edition screen is displayed. The
blinking "
" (cursor) goes out and the setting value blinks instead.
Rep. start No
S01
Rep. start No
S01
ENTER
(Selection screen)
(Edition screen)
Select the step number from which the repeat starts using the"▼▲", then press the ENTER key.
TROUBLE
ENTER
CANCEL
DOOR
Rep. start S01
S07
TROUBLE
ENTER
CANCEL
DOO
R
Rep. start S01
S07
(Decide the input) (Selection screen)
Press the CANCEL key to cancel the input. The value is cancelled and the screen returns to the
setting item selection screen.
TROUBLE
ENTER
CANCEL
DOO
R
Rep. start S03
S07
TROUBLE
ENTER
CANCEL
DOO
R
Rep. start No
S07
(Cancel the input) (Selection screen)
Press the PROGRAM key to display the next selection screen after the setting of repeat initiating
step is completed.