Manual

25
Operation Method
Input Program
1
Select program function
Check that the power in turned on.
PROGRAM
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FAN
FUNCTION
PROGRAM
MENU ENTER
CANCEL
SUBMENU
POWER
TROUBLE
Step
S01
Start in
30min
DOO
R
Press the PROGRAM key. The program menu
starts and step number selection screen is
displayed.
Press the "▼▲". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.
The "S01" is displayed when no program is
registered. In this case, the "▼▲" are invalid.
Press the ENTER key. The selected step
number is decided and the setting item selection
screen is displayed. The "Set TEMP" is
displayed first.
Press the CANCEL key to cancel the program
menu.
2
Edit step
After the step number is decided, the setting item selection screen is displayed. Select the items
on the Sub Display using the PROGRAM key.
No. Item Sub Display Notes
1
Setting
temperature
Set TEMP
100
S01
Settable value: 0 to 260
(270 maximum)
Use with the temperature of 260 or less
regardless of the setting temperature range.
2 Setting time
Set time 0min
S01
Settable value: 0 min to 59min, 1h00min to
999h59min, or "End"
3
Rotation speed
of fan
Fan 10
S01
Settable value: 1 to 10, OFF
4
Repeat
initiating step
Rep. start S01
S01
Input the step number of repeat initiating step
or "No".
5 Repeat count
Rep. count 5
S01
Settable value: 1 to 9999, "Endless"
6 Step insertion
Insert step
S01
Add a new step at the position of step
currently referred to. The sequence number
of each step hereafter increases by one.