Manual
45
Specification
Product Name Coolline
Model CLS301 CLS400 CLS600
Circulation unit Circulation in the external open system
Usable ambient temp.
5 to 30℃
Temperature control range
-10℃ to Room temperature
Temperature setting range
-15 to 35℃
Temperature adjustment
accuracy
±1.5 to 2.0℃
Refrigerator
Approx.450W at15℃ Approx.570W at15℃ Approx.820W at15℃
Maximum flow rate of pump
(50/60 Hz)
10/11 L/min 15/17 L/min
Performance
Maximum head of pump
(50/60 Hz)
4.9/6.9m 8/11m
Bath SUS304
Temperature control method Refrigerator ON-OFF control
Sensor T-thermocouple
Temperature setting method Digital setting by up/down keys
Display method Digital display
Air-cooled rotary
Refrigerator
200W 350W 600W
Cooling medium HFC R404A 270g HFC R404A 370g HFC R404A 475g
Cooling coil Copper
Outside diameter of
external circulating nozzle
Both discharge and return ports: φ13 hose nipple
Configurations
Circulation pump Magnet pump 20W Magnet pump 45W
Safety devices
Over current earth leakage breaker, Refrigerator overload relay protecting
circuit, Pump thermal protector (Pump built-in), Refrigerator pressure
detection, Delay timer for refrigerator protection, Bypass for circulating
pump protection, Dustproof filter for condenser, Key lock Function
Other functions Operation monitor, Drain cock, Temperature output terminal
Bath dimensions
(Inner dia.×height)
φ120×200 mm φ150×200 mm
External dimensions
(W×D×H)
380×460×500 mm 380×460×720 mm 380×565×720 mm
Bath capacity 1.5L 3L
Power supply (50/60Hz) 100V AC, 4A 100V AC, 6A 100V AC, 10A
Standard
Weight Approx. 40kg Approx. 45kg Approx. 60kg
Optional accessories
1-meter-long insulation hose: 1, 1-meter-long insulation hose with a
priming pump: 1, Drain hose 0.5m: 1, Wire clamp: 4, Instruction manual