Manual

31
Operation Method
Temperature Output Terminal
Specification
Temperature Output
(ANALOG)
The voltage (DC) corresponding to the measured temperature is output.
Output temperature range: CLS301 -15 to 35, CLS400/600 -20 to 35
Output voltage: 4 to 20A DC
Load: 600 or bellow
Resolution: ±1
Connection: M4 screw terminal block
CLS301
Temperature Output
-15
-2.5
10
22.5
35
4 8 12 16 20
Output Voltage (mA)
Temp.
CLS400/600
Temperature Output
-20
-6.25
7.5
21.25
35
4 8 12 16 20
Output Voltage (mA)
Temp.