Manual

21
Operation Method
Auto Stop Operation
3. Select auto stop operation
Press TIMER key, and display the character "AStP" which means
auto stop operation.
Measured temperature screen:
"AStP" which means auto stop operation is displayed.
Setting temperature screen:
The time which is set just before is displayed.
4. Set the timer
Setting value will blink at setting temperature screen.
Set the proper time by ▼▲ key.
The setting value will be smaller by key, and larger by
key.
Timer function:
The maximum setting time is "999 hours and 50 minutes".
The time can be set in increments of a minute under 99 hours
and 59 minutes.
It can be set in increment of ten minutes over 100 hours.
The "▼▲" can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment
is needed.
5. Start timer operation
Press RUN/STOP key for a second.
AUTO STOP lamp blinks and the operation will start.
Timer operation starts when the inner bath temperature at the
measured temperature screen reaches the setting temperature.
During timer operation, the remaining time is displayed at the
setting temperature screen.