Manual

19
Operation Method
Quick Auto Stop Operation
4. Stop/terminate timer operation
The operation will be stopped automatically at the setting time.
The character "End" which tells that the operation is ended will
blink at the setting temperature screen, while FIXED TEMP. lamp
and AUTO STOP lamp are on.
End the timer operation mode by pressing RUN/STOP key for a
second.
The screen will return to initial setting screen.
5. To suspend quick auto stop operation
End timer operation mode by pressing RUN/STOP key for a
second.
The screen will return to initial setting screen.
To change the setting time…
To change the setting time during the operation, press TIMER key and set the proper time by ▼▲ key.
In this case, it is necessary to add the value of elapsed time to newly adding time. After a while, the blink
at the setting temperature screen will stop, and the change will be confirmed.