Owner manual

51
In the Event of Failure…
Safety Device and Error Code
This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The table below shows the cause and the solution method when the safety device operates.
Error Code:
When an abnormal condition occurs, an error code appears and the alarm lamp lights in the
controller, the buzzer sounds simultaneously. Record the error code and turn off the power of
device immediately.
Safety Device Notify Cause/Solution
Sensor trouble detection
ALARM” lamp lights on,
Er.01” appears
Temperature sensor is broken or
disconnected.
Make a call for service.
SSR short-circuit detection
ALARM” lamp lights on,
Er.02” appears
Triac is in short-circuit
Make a call for service.
Heater disconnecting
detection
ALARM” lamp lights on,
Er.03” appears
Heater is disconnected.
Make a call for service.
Measured temperature lower
limit error
ALARM” lamp lights on,
Er.13” appears
When the temperature warning
function issues a lower limit warning
Make a call for service.
Memory error
ALARM” lamp lights on,
Er.15” appears
Failure in internal memory.
Make a call for service.
Internal communication error
ALARM” lamp lights on,
Er.17” appears
Failure in internal communication or
temperature inputting circuit.
Make a call for service.
Overheating
ALARM” lamp lights on,
Er.19” appears
Overheating prevention device is in
operation.
Reset the power supply, and then
adjust the setting temperature of the
overheating protection device.
If the state does not recover, make a
call for service.
Flow rate error
ALARM” lamp lights on,
Er.20” appears
The circulating fluid does not properly
circulate.
Air remains in the circulating path.
Measurement temperature
error
ALARM” lamp lights on,
----” appears
Measurement value is out of display
range.
Make a call for service.
Refrigerator pressure error
REFRIGERATO ERROR
lamp lights on
The condenser filter is dirty.
The room temperature is high.
The temperature of the circulating fluid
is 40°C or higher.