Operation Method
Temperature Output Terminal
Specification
Temperature Output
(ANALOG)
• The voltage (DC) corresponding to the measured temperature is output.
• Output temperature range: CLH300 -5 to 85℃, CLH400/600 -20 to 85℃
• Output voltage: 4 to 20mA DC
• Load: 600Ω or bellow
• Resolution: ±1℃
• Connection: M4 screw terminal block
CLH300
Temperature Output
-15
8.75
32.5
56.25
80
4 8 12 16 20
Output Voltage (mA)
Temp.
CLH400/600
Temperature Output
-20
5
30
55
80
4 8 12 16 20
Output Voltage (mA)
Temp.