Owner manual

28
Operation Method
Temperature Output Terminal
Specification
Temperature Output
(ANALOG)
The voltage (DC) corresponding to the measured temperature is output.
Output temperature range: CLH300 -5 to 85, CLH400/600 -20 to 85
Output voltage: 4 to 20A DC
Load: 600 or bellow
Resolution: ±1
Connection: M4 screw terminal block
CLH300
Temperature Output
-15
8.75
32.5
56.25
80
4 8 12 16 20
Output Voltage (mA)
Temp.
CLH400/600
Temperature Output
-20
5
30
55
80
4 8 12 16 20
Output Voltage (mA)
Temp.