Specification
Product Name Desktop type low constant temperature water bath
Model BBL100 BBL300
Circulation unit External sealed unit circulation
Usable ambient temp.
5℃∼30
Temperature control
range
-10℃∼80
Temperature adjustment
accuracy
±0.1
Temperature distribution
accuracy
±0.3
Cooling capacity
Approx.300W at15 Approx.410W at15
Circulating/stirring
method
Magnet pump circulation cooling/external sealed unit circulation
Maximum flow rate 3.7/4.7 L/min (50/60Hz)
Performance
Maximum head 1.0/1.5 m
Bath SUS304
Temperature control
system
PID control by microcomputer
Sensor
Pt100Ω (for temperature controller) and
K-thermocouple (for overheating prevention device)
Temperature
setting/display method
Digital setting / Digital display
Overheating prevention
sensor
K-thermocouple
Stainless pipe heater SUS316L
Heater
700W 900W
Air-cooling sealed rotary
Refrigerator
150W 250W
Cooling medium R134a 310g R404A 310g
Cooling coil Copper (nickel-plated)
Configurations
External circulating
nozzle size
Discharge port and return port have outer diameter 10.5mm hose nipple
Safety devices
Earth leakage breaker, Overheating prevention device, Pump thermal
protector, Refrigerator overload relay protecting circuit, Delay timer for
refrigerator protection, Refrigerator pressure detection, Float switch,
Self-diagnostic functions (Failure of sensor, Heater disconnection, SSR
short-circuit, Automatic overheating prevention)
Other functions
Drain cock, Key lock Function, Calibration offset, Temperature output terminal,
RS485 communication function, Refrigerator pressure indicator, condenser
filter
Bath dimensions
(W×D×H)
300×150×177.5 mm 300×240×177.5 mm
Effective internal
dimensions (W×D×H)
300×150×155 mm 300×240×155 mm
External dimensions*
(W×D×H)
400×530×500 mm
(Height of bath: 330 mm)
400×600×500 mm
(Height of bath: 330 mm)
Bath capacity
(Effective capacity)
8L (6.75L) 13L (11.5L)
Power supply (50/60Hz) 100V AC 9A 100V AC 12.5A
Standard
Weight Approx.35kg Approx.55kg
Accessories Bottom shelf, Cover, 0.5m Drain hose, 0.5m overflow hose, Instruction manual