Datasheet
- 3 -
Specification
Type HV-2
Voltage 100-120Vac 220-240Vac
Frequency 60Hz 50Hz
Current 1.2A 0.6A
Rated Output Power 130W 120W
Fuse T3.15A T2A
Air Flow
340m
3
/hr
Range of Regulating Flow
80%~100%
Noise Level
<
50
dB
Outline Dimension
310(L ) × 310(W ) × 460(H) mm
Weight 20kg
Setting Up
1. Unpack your new extraction system and ensure that the working voltage is
the same as that of your electrical supply and truckles are lifted clear.
2. The inlet(s) on system are located on the top of the unit and sealed with
cap(s)
3. The unit comes with a built inside silencer as standard.
4. The filtered air is exhausted through the base. Under no circumstances
should this outlet be covered as this will restrict airflow and cause
over heating, may damage the system.
5. The unit must not be sited in an area exceeding 40℃.
6. Check filters are located in their correct position and carefully replace lid.
7. Turn on the rocker switch. The unit will take a few seconds to build up to
speed on initial start up.
8. Unit fitted with variable flow control should be adjusted to the desired flow.
Note: Unit with flow control has a 10-15 second delayed
after shut down before re-start.
Operating Instructions
*
Green LED: Operational range is 0 to 1200 hours.
*
Orange LED: Operational range is 1201 to 1300 hours.
* Red LED: Operational range is 1301 to 1439 hours.
* Red LED flashing warning and buzzer: Operational time has exceeded 1440 hours (Need to