Owner's manual

WQL-Cond What to do if...
185
ba75822e01 03/2011
13 What to do if...
13.1 Calibration and measuring
Unstable
measured values
Implausible
measured values
Incorrect
temperature
display
Logger does not
react
System cannot be
calibrated
Cause Remedy
No stable measured value Provide stable measuring
conditions (e.g. temperature)
Use a shorter USB cable
USB-Hub or USB connection
at the PC faulty or defective
Use other USB connection or PC
Cause Remedy
Measuring range exceeded Make sure the logger is suitable
for the application
Contamination in the area of
the electrodes
Clean the conductivity measuring
cell (see section 12.2 E
XTERIOR
CLEANING).
Electrodes damaged Return the logger
Cause Remedy
The temperature sensor is not
immersed deep enough in the
measuring solution
Observe the minimum immersion
depth
Temperature sensor defective Return the logger
Cause Remedy
Battery depleted Insert new battery (see section
12.3)
Cause Remedy
Contamination in the area of
the electrodes
Clean the conductivity measuring
cell (see section 12.2 E
XTERIOR
CLEANING).
Calibration solution not suitable Use suitable control standard