Owner's manual

WQL-Cond Calibration in control standard (determining the cell constant)
133
ba75822e01 03/2011
Note
You can prematurely terminate the AutoRead function manually with Accept
at any time. If the AutoRead function is prematurely terminated, the current
measurement data are accepted immediately.
5 Use the Continue button to start the measurement.
6 The note, AutoRead is running... appears.
As soon as the value is stable, the prompt to enter the temperature
dependent conductivity value appears.