WQL-Cond Calibration in control standard (determining the cell constant)
129
ba75822e01 03/2011
6.2 Calibrating (automatic)
Preparatory
activities
Calibration
Note
We recommend to always calibrate with the protective hood screwed on so
the measuring cell is protected. Use a stand as necessary.
1 Connect the WQL-Cond logger to a USB port of your PC with the
USB cable.
2 Have the 0.01 mol/l KCl control standard ready.
3 In the Device menu, click the menu item, Calibrate.
The calibration window appears.
4 Immerse the logger in the 0.01 mol/l KCl control standard.
5 Use the Next button to start the measurement.