User guide

86GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Sensor Error - General sensor error that may also
be transferred to a relay output. Contact
MJK.
Sensor communication Error - Modbus communication errrors nor-
mally imply that a device has been moved,
added, removed or has changed status in an
illegal manner. Rearranged cables or wires
can also invoke this error message.
Reset the system to cancel the error condi-
tion.
Sensor hardware Error - Change the sensor hardware.
Sensor setting Error - Some settings are wrong or have
been changed.
Sensor supply Error - The sensor supply is too low or too
high. Check the sensor cable connection.
System Error - Internal VCC error. See the event log
for details.
System error Error - The system software is inoperable in-
cluding faulty alarm and sensor bits. Contact
MJK.
D.O. offset Error - A zero-point sensor problem. The
sensor needs a zero-point calibration.
Contact MJK.
Warnings
Sensor range exceeded Warning - The measured value is out of
range. Check the measuring range and the
measuring units.
Info
Measurement hold Info - The measurement is set on hold dur-
ing the cleaning cycle or by an error.
Sensor needs cleaning Info - The sensor needs cleaning.
System reset occurred Info - The converter has been subject to a
software or hardware reset.