Dissolved Oxygen Transmitter
GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
The data log contains:
Date•
Time•
Dissolved oxygen values •
In case of a power failure, the data logger continues when power returns.
If more converters are connected to one Display Unit, each converter has its
own individual log interval and can be sorted.
All converters share the same memory of 20.000 entry points.
The log data can be shown on the Display Unit or be stored in an external
CSV file. MJK-Field Link software is needed for transfer of data into CSV file
format via the USB port in the display unit. The format can be read in for ex-
ample Microsoft
®
Excel
®
(see Appendix D on page 97 for details).
An example of the log capacity of one sensor versus the time interval is
shown in the following table.
Log interval Log capacity
Seconds Minutes Hours Days
10 - 55,56 2,31
30 - 166,67 6,94
- 1 333 14
- 5 1667 69
- 10 3333 139
- 30 10000 417
- 60 20000 833