User guide

30GB Oxix Manual 080915
FW: 841013-009, 846001-002, 841513-009
Dissolved Oxygen Transmitter
Sensor and Mounting Kit / Connection Board Terminals
Important!: Make sure the right wiring indication label is fitted below
the terminal block (label indicated by arrow). The label
must clearly be marked Oxix
®
.
Wrong wiring may cause sensor destruction !
Connect the sensor wires to the Oxix
®
converter with the wires coming from
the sensor as shown below (see arrow).
Mains
supply
-mA- -D02--D01--DI--Display-
+ +A B C2 N02C NO N01 PE N L
- -
C1
Line
2 wire
shielded cable.
4 wire
shielded cable.
display unit
display unit
Display unitConverter #4Converter #1
Max. four MJK converters with Modbus
can be connected to one display unit:
521274
2,5
Main supply
Supply Fuse
10-30 V DC 1 A T
24 V AC 500 mA T
115 V AC 125 mA T
230 V AC 63 mA T
Sensor supply
Fuse 200 mA T
+DCGND A B
Red
Black
White
Green
Shield
521275
www.mjk.com
DK: +45 45 56 06 56
NO: +47 69 20 60 70
SE: +46 53 31 77 50
NL: +31 251 672171
USA: +1 847 482 8655
AUS: +61 3 9758 8533
Oxix
®
converter terminals
Terminals
Wires from sensor Converter terminals
- not used
Black GND
White A
Green B
Shield - (shield GND)
Red +DC