SDI-12 Command and Response Protocol 5-17
Example of a H-3521 Extended "Read Mean_Count" command:
Command Response Time Values Description
"aXRMC!" "a0011<cr><lf>" 1 sec 1 Read Mean_Count
Command Response Description
"aD0!" "a+8<cr><lf>" Count = 8
Example of a H-3521 Extended "Write Mean_Count" command:
Command Response Time Values Description
"aXWMC4!" "a0021<cr><lf>" 2 sec 1 Write Mean_Count
Command Response Description
"aD0!" "a+4<cr><lf>" Count = 4