User's Manual

6
Menu Menu
Note: If the matching refers to clamp oximeter is basic, not equipped with parameter setting
function
Detailed descriptions of product functions:
1. Display TypeOLED display
2. SpO2Measurement range70%100%
Accuracy80%100%:±2% (Including 80%);
70%79%: ±3% ;
Below 70% no requirement;
Resolution1%
3. PRMeasurement range30BPM254BPM
Accuracy≤100BPM ,±1BPM
100BPM ,±2BPM
4. Parameters of LED probe
wavelength
radiation power
RED
660±5nm
1.8mW
IR
940±10nm
2.0mW
This parameter can be especially useful to clinicians.
5. PI indexmeasurement scope0~20 optional
6. Power two AAA 1.5V alkaline batteries
7. Automatic standbythe product shuts off by itself when no finger is in the product about 8
seconds
8. Automatic start up: every 5 s instrument will automatically detect the signal, after the hole with
my finger, timely automatically boot;(optional)
9. Gravity sensing function: finger movement, the screen display will change with the gravity
sensing changes (optional)
10. Dimension58mm×36mm×33mm
11. Operation environment:
Temperature: 5 ~ 40
Humidity: 15% ~ 80%,no condensing
Atmospheric pressure: 70kPa ~ 106kPa
Transport, storage environment:
Temperature: -10 ~ 40
Humidity: 10%~ 95%,no condensing
Atmospheric pressure: 70kPa ~ 106kPa, non-corrosive gas and well-ventilated
environment.
12. DeclarationEMC of this product comply with IEC60601-1-2 standard.
13. This equipment has no alarm function.
14. Data averaging and signal processing delay the display and transmission of SpO2 data.
The date update period :12S
15. About 30min required for oximeter to warm from the minimum storage temperature between
uses until it is ready for intended use