Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30A/m
Guidance and manufacturer´s declaration electromagnetic immunity
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer´s declaration electromagnetic immunity
The Fingertip Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Fingertip Pulse Oximeter should assure that
it is used in such an environment.
Immunity test
EN 60601 test level
Compliance level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
N/A
Table 9 - Test specifications for enclosure port immunity to RF wireless
communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385
380 -390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0.3
9
745
780