Hardware requirements:
Noraxon Hardware
Xsens Hardware
Wireless EMG transmitters
1 TeleMyo DTS (plus antenna)
1 TeleMyo mini-receiver (plus antenna)
MVN System
Awinda Station
Both systems of course also need related cables to connect to each other (with BNC connectors at each
end) and a USB cable to connect to the PC.
In addition to the normal MTw hardware setup, set up the hardware of the Noraxon system as follows:
USB port of TeleMyo mini-receiver to USB of PC.
Connect jack connector to Sync Out port of TeleMyo mini-receiver to BNC connection Sync In 1 of
Awinda Station.
Manual trigger pulse, jack connector to Sync In port of TeleMyo mini-receiver.
Connect the external antenna to the TeleMyo DTS.
When successfully connected and switched on, the TeleMyo DTS will display “WiFi ready”.
23.7.3.5 Software Setup: MVN Studio
Sync In
Select Start Recording
Select In 1
Polarity: Rising Edge
Trigger Once: Uncheck
Skip first = 0
Skip factor = 1
Select Stop Recording
Select In 1
Polarity: Rising Edge
Trigger Once: Uncheck
Skip First = 1
Skip Factor = 1
To initialize recording, ensure that the MVN System has been correctly initialized (body dimensions and
well calibrated). MVN Studio will wait for a signal from the Awinda Station to start a recording.
23.7.3.6 Software Setup: Noraxon MyoResearch Software:
Based on the output settings described for MVN Studio, the settings for Noraxon MyoResearch software
can remain the same. The difference is that instead of the trigger pulse coming from the record button
in MVN Studio, this now comes from the manual button connected to the mini-receiver. Additionally, the
mini-receiver should be set up as follows:
Go to the hardware menu;
Select the TeleMyo mini-receiver from the list of icons;
Select: settings;
Select: Configure;
Ensure that the wireless sync is “External Pulse” and Input Range is ±5V