User Manual
© Xsens Technologies B.V.
76
Manual trigger pulse, jack connector to Sync In port of TeleMyo minirecevier.
Connect the external antenna to the TeleMyo DTS.
When successfully connected and switched on, the TeleMyo DTS will display “WiFi ready”.
11.8.5.5 Software Setup: MT Manager
Synchronisation on the Sync In Line 1 port:
Sync In
Select Start Recording
Check the check box for Line 1
Polarity: Rising Edge
Trigger Once: Uncheck
Skip first = 0
Skip factor = 1
Select Stop Recording
Check the check box for Line 1
Polarity: Rising Edge
Trigger Once: Uncheck
Skip First = 1
Skip Factor = 1
Set up the wireless configuration in MT Manager.
To initialise recording, click the Record button. To indicate that MT Manager is waiting for an
external pulse, the icon changes from the normal red dot to one with the pause symbol
overlaid, as shown in Section 6.7.2.
11.8.5.6 Software Setup: Noraxon MyoResearch Software:
Based on the output settings described for MT Manager, the settings for Noraxon
MyoResearch software can remain the same. The difference is that instead of the trigger
pulse coming from the record button in MT Manager, this now comes from the manual
button connected to the mini-receiver. Additionally, the mini-receiver should be set up as
follows:
Go to the hardware menu;
Select the TeleMyo mini-receiver from the list of icons;
Select: settings;
Select: Configure;
Ensure that the wireless sync is “External Pulse” and Input Range is ±5V