Product Information

21
Medical Equipment Safety
This portable equipment and mobile RF communications equipment can affect medical
electrical (ME) equipment.
This medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service per the EMC information provided in the accompanying
documents.
This ME equipment or ME system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the ME equipment or ME system
should be observed to verify normal operation in the configuration in which it will be used.
The use of accessories, transducers, and cables other than those specified, except for transducers and
cables sold by the manufacturer of the equipment or system as replacement parts for internal
components, may result in increased emissions or decreased immunity of the ME equipment or ME
system.
Table 1-1: Guidance and manufacturer’s declaration electromagnetic emissions for all ME equipment and ME
Guidance and manufacturer’s declaration electromagnetic emissions
The XSLATE L10 and XPAD L10 Rugged Tablet PC is intended for use in the electromagnetic environment specified below. The customer or
the user of the XSLATE L10 Rugged Tablet PC should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions (CISPR 11)
Group 1
The XSLATE L10 Rugged Tablet PC uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The XSLATE L10 Rugged Tablet PC is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
RF emissions (CISPR 11)
Class B
Harmonic emissions (IEC 61000-
3-2)
Class A
Voltage fluctuations/ flicker
emissions (IEC 61000-3-3)
Complies
Notes:
None
Table 1-2: Guidance and manufacturer’s declaration electromagnetic immunity For all ME equipment and ME systems
The XSLATE L10 and XPAD L10 Rugged Tablet PC is intended for use in the electromagnetic environment specified below. The customer or
the user of the XSLATE L10 and XPAD L10 Rugged Tablet PC should assure that it is used in such an environment.
Immunity test
Test level
Compliance level
Electromagnetic environment guidance
Electrostatic discharge
(ESD) (IEC 61000-4-2)
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst (IEC 61000-
4-4)
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge (IEC 61000-4-5)
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines (IEC 61000-4-
11)
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the XSLATE L10 Rugged Tablet PC requires continued
operation during power mains interruptions, it is
recommended that the XSLATE L10 Rugged Tablet PC
be powered from an uninterruptible power supply or
a battery.