Installation Guide
CALIBRATION
39
Calibrating with the DTP41
Using the DTP41 spectrophotometer, you can measure color patches generated by the digital
press and automatically download the measurements to the Fiery EXP8000/50.
NOTE: Before you use ColorWise Pro Tools to calibrate the Fiery EXP8000/50, set up and
calibrate the DTP41. For instructions, see the documentation that accompanies the
instrument.
TO CALIBRATE THE FIERY EXP8000/50 USING THE DTP41
1 Start Calibrator.
For instructions, see page 33.
2 Choose X-Rite DTP41 as the measurement method.
3 In the Check Print Settings pane, choose a calibration set.
If more than one option appears, choose the appropriate calibration set for the type of media
and screen you use most often.
NOTE: For this calibration to take effect, the calibration set must be associated with one or
more output profiles. The default calibration set is already associated with the default output
profile. You do not need to make any new associations.
4 Click Print in the Generate Measurement Page pane.
The Print Options dialog box appears.
5 Choose the appropriate options and click Print.
Page Type: Choose 21 or 34 Sorted Patches.
Paper Size: Automatically set to LTR/A4 for 21 Sorted Patches. Choose 11x17/A3 or
12x18/SRA3 for 34 Sorted Patches.
Input Tray: Choose the paper source for printing the measurement page.
6 Retrieve the measurement page from the digital press.
7 In the Get Measurements pane, click Measure.
The Measurement Options dialog box appears, displaying the specified Page Type and Paper
Size options.