User guide

PLATESCOPE
6
Charging the Batteries
Your new PlateScope comes in a discharged condition and must be charged before use (up to 4
hours for full charge). Upon initial use (or after a prolonged storage period), the batteries (X-Rite
P/N SE15-38) may require three to four charge/discharge cycles before achieving maximum
capacity. It is normal for a battery to become warm during charging and discharging.
It is important to condition (fully discharge and then fully charge) the battery every two to three
months. Failure to do so may significantly shorten the battery's life. To discharge, simply operate
your instrument under the battery's power until it shuts down or until you get a low battery
warning (flashing battery indicator gauge). Then recharge the batteries for up to 4 hours.
Charged batteries will eventually lose their charge if unused. It may therefore be necessary to
recharge the batteries after a storage period.
The AC adapter (X-Rite P/N SE30-77L) overrides any current battery condition. Measurements
can be taken even with a very low battery condition when using the AC adapter. The instrument
can also operate using only the AC adapter with the batteries removed.
Attaching the AC Adapter
1. Verify the voltage indicted on the AC adapter complies with the AC line voltage in your area.
2. Insert the small plug from the AC adapter into the input connector on the instrument.
3. Plug the detachable line cord in the AC adapter and plug the line cord into the wall receptacle.
Operational hazard exists if an AC adapter other than X-Rite SE30-77L is used. Use only
X-Rite battery pack SE15-38; other types may burst causing personal injury.
AC adapter
AC adapter input
Detachable line cord
AC Adapter Ratings
Input: 100-240V 50-60 Hz
Output: 12VDC @ 2.5A