Please note that the validation of a proof based on an OBC-ICC profile might fail even though the
quality of the hardcopy proof is good when measured with a NoFilter spectrophotometer like the
i1Pro, as measurements from this type of device do not correct for the optical brightening agents
in the paper. If applied correctly, OBC profiles will produce better visual agreement between
hardcopy and softcopy proofs and prints. As the OBC Module creates a new group of
measurement conditions taking the optical brightening agents and the amount of UV of the
viewing condition into account, different OBC conditions such as NoFilter and UVcut Filter
measurements should not compared against each other.