9. Process beam
This process beam should help you in orientation cases to see in which
stadium the process is situated at the moment.
It’s divided in:
▪ Communication
(device to software, software to server)
▪ Repeatability
(each 10 measurements about blue and about white)
▪ Accuracy
(comparison current measurement to server values)
▪ Confirmation
(metrological check of the profile)
▪ Conclusion
(creation of certificate)
10. Reentry of temperature and measurement
To check the quality of the calculated profile the temperature has to be reentered
again. Please don’t leave too much time pass by between the first and the second
measurement to keep the parameter as similar as possible.
Please note that the positioning of the second measurement should be as exactly
as the first one. Only like this the accuracy of the calculated profile adjustment is
warranted.
If the positioning of the two measurement acts varies too much, the result after
the profiling may be worse than before.