Technical data
Settings
............................
5.6.5
5.6.5 Check calibration with absolute white base
Procedure:
........................................................................................................................................
Select absolute white base Setting k Measurement parameters k White base.
........................................................................................................................................
• Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
• Select the menu line Check calibration by turning the <Control ball>.
• Activate the selected function by pressing the <Measurement key>.
• Position the measuring aperture on the white field of the calibration card and initiate a measurement
by pressing the <Measurement key>.
The display shows White measurement done for about 1 second.
• Then the request Measure color is shown on the display.
• Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>.
• The calibration values and the corresponding colors appear on the display one after the other.
• Compare the calibration values with the target values on the calibration card.
If the values agree or if the difference is not more than ± 0.02 D, then it is not necessary to
recalibrate the densitometer.
• The sign
✓ behind the calibration values on the display indicates that the measured values are
within the specified tolerance.
• • • • •