Technical data

Settings
............................
5.6.5
5.6.5 Check calibration with absolute white base
Procedure:
........................................................................................................................................
Select absolute white base Setting k Measurement parameters k White base.
........................................................................................................................................
Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Select the menu line Check calibration by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
Position the measuring aperture on the white field of the calibration card and initiate a measurement
by pressing the <Measurement key>.
The display shows White measurement done for about 1 second.
Then the request Measure color is shown on the display.
Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>.
The calibration values and the corresponding colors appear on the display one after the other.
Compare the calibration values with the target values on the calibration card.
If the values agree or if the difference is not more than ± 0.02 D, then it is not necessary to
recalibrate the densitometer.
The sign
behind the calibration values on the display indicates that the measured values are
within the specified tolerance.
• • • • •