Technical data

Settings
............................
5.6.3
Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>.
The calibration values and the corresponding colors appear on the display one after the other.
Compare the calibration values with the target values on the calibration card.
If the values agree or if the difference is not more than ± 0.02 D, it is not necessary to
recalibrate the densitometer.
The sign
behind the calibration values on the display indicates that the measured values
are within the specified tolerance.
After the color Yellow has been measured, and if all calibration values are within the specified
tolerance, the graphics bar will jump to the menu line Return.
Press the <Measurement key> several times to return to the main menu.
• • • • •