Technical data
Settings
............................
5.6.3
• Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>.
• The calibration values and the corresponding colors appear on the display one after the other.
• Compare the calibration values with the target values on the calibration card.
If the values agree or if the difference is not more than ± 0.02 D, it is not necessary to
recalibrate the densitometer.
• The sign
✓ behind the calibration values on the display indicates that the measured values
are within the specified tolerance.
• After the color Yellow has been measured, and if all calibration values are within the specified
tolerance, the graphics bar will jump to the menu line Return.
• Press the <Measurement key> several times to return to the main menu.
• • • • •