Technical data
Settings
............................
5.6.2 – 5.6.2.1
• Press the <Measurement key> to get into the entry field of the color Black.
• Turn the <Control ball> to correct the calibration value until it corresponds with the target value
indicated on the calibration card.
• Terminate the entry by pressing the <Measurement key>.
• Correct the calibration values of the colors Cyan, Magenta and Yellow one after the other.
• The graphics bar jumps to the menu line Accept data.
• Store the calibration data by pressing the <Measurement key>.
• The display shows Check calibration in reverse video.
• • • • •
5.6.2.1 Reject data
Procedure:
• Select the menu line Reject data by turning the <Control ball> and confirm this action by pressing
the <Measurement key>.
The new data will not be stored; previously stored data are retained.
• The menu lines Return, Check calibration and New calibration appear on the display.
• Select the desired menu line and confirm the selection by pressing the <Measurement key>.