Technical data
5
Settings
............................
5.6.2
5.6.2 Calibration with paper white base
Procedure:
........................................................................................................................................
Select paper white base under Settings k Measurement parameters k White base.
........................................................................................................................................
• Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
• Select the menu line New calibration by turning the <Control ball>.
• Activate the selected function by pressing the <Measurement key>.
• Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
• The display shows White measurement done for about 1 second.
• Then the request Measure color is shown on the display.
• Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>. The measurements can be made in any order.
• The uncorrected calibration values and the corresponding colors are shown on the display one after
the other.
• After the last measurement, the graphics bar will jump to the color Black.
• • • • •