Technical data

5
Settings
............................
5.6.2
5.6.2 Calibration with paper white base
Procedure:
........................................................................................................................................
Select paper white base under Settings k Measurement parameters k White base.
........................................................................................................................................
Select the menu line Calibration by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Select the menu line New calibration by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display shows White measurement done for about 1 second.
Then the request Measure color is shown on the display.
Position the measuring aperture on the color fields to be measured and initiate a measurement each
by pressing the <Measurement key>. The measurements can be made in any order.
The uncorrected calibration values and the corresponding colors are shown on the display one after
the other.
After the last measurement, the graphics bar will jump to the color Black.
• • • • •