Technical data

5
Settings
............................
5.4.5
5.4.5 Reference entry
Procedure:
Select the menu line Reference entry by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
Activate the function by pressing the <Measurement key>.
The display shows Measure paper white in reverse video.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display White measurement done is shown for about 1 second.
The display shows Measure color in reverse video.
Position the measuring aperture on the color samples Black, Cyan, Magenta and Yellow, and initiate
the measurements by pressing the <Measurement key>.
If more than four references have been defined for this job, a table with a proposal for the
assignment of the reference will be shown in reverse video after every measurement.
Select the respective reference by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.