Technical data
Densitometric functions
............................
4.11
4.11 Trapping
........................................................................................................................................
Select paper white or absolute white base under Settings k Measurement parameters k
White base.
........................................................................................................................................
........................................................................................................................................
Select automatic or manual filter selection under Settings k Color or in the selected
densitometric function under the menu line Color.
........................................................................................................................................
Procedure:
• Select the menu line Trapping by turning the <Control ball>.
• Activate the selected function by pressing the <Measurement key>.
• Select the menu line Measure paper white by turning the <Control ball>.
• Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
• The display White measurement done is shown for about 1 second.
• Position the measuring aperture on the solid tone patch printed first and initiate a measurement by
pressing the <Measurement key>.
• The density value of the first solid tone patch is shown on the display.
• Position the measuring aperture on the solid tone patch printed second and initiate a measurement
by pressing the <Measurement key>.
• The density value of the second solid tone patch is shown on the display.
• Position the measuring aperture on the overprint and initiate a measurement by pressing the
<Measurement key>.
• The overprint value T is shown on the display.
• Select the menu line Return by turning the <Control ball> and press the <Measurement key> to
return to the main menu.