Technical data
Densitometric functions
............................
4
4.6
4.6 Dot gain
........................................................................................................................................
Select automatic or manual filter selection under Settings k Color or in the selected
densitometric function under the menu line Color.
........................................................................................................................................
........................................................................................................................................
Enter the measurement parameters for the dot gain under Settings k Measurement
parameters k Dot gain. One, two or three film reference values may be entered.
........................................................................................................................................
Procedure:
• Select the menu line Dot gain by turning the <Control ball>.
• Activate the selected function by pressing the <Measurement key>.
• Select the menu line Measure paper white by turning the <Control ball>.
• Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
• The display White measurement done is shown for about 1 second.
• Position the measuring aperture on the solid tone patch and initiate a measurement by pressing the
<Measurement key>.
• The density value of the solid tone patch is shown on the display.