Technical data

Appendices
............................
11
Appendix B
1 Serial interface
All D190 remission densitometers are equipped with a true serial RS232 interface with bipolar levels.
1.1 Specifications
1.2 Parameters
1.3 Pin assignment of Mini DIN socket
1 HS_POWER
2 HS_OUT
3 DATA_IN
4 GND
5 DATA_OUT
6 HS_IN
7 CRG+
8 CRG-
B 1 – 1.3
21
35
4
68
7
Parameter Specification Remarks
Data transmission rate 110 ... 19200 baud
Data bits 8
Parity bit no parity bit
Stop bits 1
Handshake no „handshake“
software handshake: Xon/Xoff Xon = ASCII (17), Xoff = ASCII (19)
hardware handshake see 1.4
Description D190 input D190 output
Level for logic 0 -15.0 V ...+ 0.5 V - 8.0 V ... - 5.0 V at I
L
<1.6 mA
Level for logic 1 + 3.5 V ...+15.0 V +8.0 V ... +5.0 V at I
L
<1.6 mA
Input resistance > 6 kOhm