Technical data
Maintenance
............................
8
........................................................................................................................................
The D190 remission densitometer can be operated either with a nickel-cadmium or a
metal-hydride battery.
The battery type is coded on the connector and will be registered by the software.
........................................................................................................................................
........................................................................................................................................
Use only original GretagMacbeth accessories and spare parts.
........................................................................................................................................
8.1 Battery replacement
The battery can be replaced without tools by the user.
Procedure:
• Open battery compartment cover (snap connection).
• Remove battery and disconnect plug.
• Plug in cable of new battery (coded plug).
• Insert new battery in compartment and close cover.
8.2 Adjustment of the measuring aperture
Procedure:
• Unscrew mounting screws of measuring aperture until aperture can be
moved.
• Manually set adjustment knob to position A.
• Withdraw measuring module manually until adjustment knob is in
position B (matched arrows).
• Insert white paper between measuring module and measuring aperture.
• Switch on measuring lamp: Settings
k Device settings k Measure
system
k Lamp <on>.
• Center measuring aperture on light spot.
• Retighten mounting screws.
8.1 – 8.2