Technical data

Maintenance
............................
8
........................................................................................................................................
The D190 remission densitometer can be operated either with a nickel-cadmium or a
metal-hydride battery.
The battery type is coded on the connector and will be registered by the software.
........................................................................................................................................
........................................................................................................................................
Use only original GretagMacbeth accessories and spare parts.
........................................................................................................................................
8.1 Battery replacement
The battery can be replaced without tools by the user.
Procedure:
Open battery compartment cover (snap connection).
Remove battery and disconnect plug.
Plug in cable of new battery (coded plug).
Insert new battery in compartment and close cover.
8.2 Adjustment of the measuring aperture
Procedure:
Unscrew mounting screws of measuring aperture until aperture can be
moved.
Manually set adjustment knob to position A.
Withdraw measuring module manually until adjustment knob is in
position B (matched arrows).
Insert white paper between measuring module and measuring aperture.
Switch on measuring lamp: Settings
k Device settings k Measure
system
k Lamp <on>.
Center measuring aperture on light spot.
Retighten mounting screws.
8.1 – 8.2
U