Technical data
7
Help
............................
Message Message on Error source
No. display
(34) Charger connected A charger is connected to
the D190
(35) Charger disconnected The charger was
disconnected before the
battery was fully charged
39 Charge time-out The charging procedure
was not duly completed
43 A/D failure The values from the
analogue to digital
converter are wrong
45 Drive error The filter wheel position
pulses are missing
46 Function released The password was
entered correctly and the
new function is now
available.
47 Password wrong A wrong password has
been entered.
48 Trapping not available In the Auto function an
overprint field has been
measured without having
the trapping function
49 Dot gain not available In the Auto function a
halftone field has been
measured without having
the dot gain function
7.7
Reméde
Connect charger again to
fully charge the battery
Check battery and replace
it if necessary
Please contact your
GretagMacbeth repre-
sentative
Please contact your
GretagMacbeth repre-
sentative
Check entry. If the
password is OK, please
contact your Gretag-
Macbeth representative
If you wish to measure
overprint fields, you can
by the function. Contact
your GretagMacbeth
representative
If you wish to measure
halftone fields, you can by
the function. Contact your
GretagMacbeth repre-
sentative