Technical data
7
Help
............................
Message Message on Error source
No. display
3 Motor blocked Drive of measuring
module blocked or
defective
4 Lamp failure Measuring lamp defective
7 No printer connected The printer does not send
a reply to the query of the
D190
9 Measure locked The measurements were
disabled by remote
control commands
13 EPROM failure EEPROM defective
22 Division by zero A division by zero was
detected during the
calculation.
7.7
Actions
If no further measure-
ments can be performed
please contact your
GretagMacbeth repre-
sentative
Exchange measuring lamp
Connect printer Check
settings of printer and
D190
Enable measurements by
remote control commands
Please contact your
GretagMacbeth
representative to
exchange the EEPROM
Check measurement
sequence and measure-
ment results. Insert the
measurement results in
the corresponding formula
given in appendix A.
Select the measurement
values in a way that no
division by zero occurs.
7.7 Messages
........................................................................................................................................
Messages are shown on the display as follows:
Message: (1...n)
Brief description
Message numbers in brackets are not shown on the display.
........................................................................................................................................