Technical data
6
Device settings
............................
6.9.1.3 – 6.9.2
6.9.1.3 Measurement counter
The measurement counter indicates the number of performed measurements.
........................................................................................................................................
The measurement counter is reset to zero with the function Initialize device .
See Settings
k Device settings k Initialize device .
........................................................................................................................................
6.9.2 Initialize device
........................................................................................................................................
The D190 densitometer can be reset to factory-stored basic values with the function Initialize
device.
The user defined values will be replaced by default values!
........................................................................................................................................
........................................................................................................................................
The function Initialize device includes:
• Deleting user-defined functions and switching on all functions
• Setting the manual color selection to automatic color recognition
• Deleting user-defined orders and activating the basic setting
• Deleting user-defined reference and tolerance values and setting basic values
• Deleting user-defined parameters and setting basic values
• Deleting calibration values for paper white and absolute white base
• Changing the Device settings to basic values
........................................................................................................................................
Procedure:
• Select the menu line Initialize device by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
• The display shows Return in reverse video.
• Select the menu line Initialize device by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
• The display shows Delete all data ? No.
• Turn the <Control ball> to select No or Yes.
Yes: • After pressing the <Measurement key>, the unit will be re-initialized and the display
deleted.
No: • The display shows Return in reverse video.
• After repeatedly pressing the <Measurement key>, the program will return to the main menu.