Technical data

Device settings
............................
6.9.1.1 – 6.9.1.2
Dark pulse number
12000 ... 24000 typ.18000
Color Measuring range Channel sensitivities
B, C, M, Y 0.00 D ... 2.50 D 0.00 ... 0.70 typ. 0.30
Y (NB) 0.00 D ... 2.20 D 0.00 ... 1.00 typ. 0.60
6.9.1.1 Switching the lamp on/off
Procedure:
Select the menu line Lamp by turning the <Control ball>.
Press the <Measurement key>; the display shows <off>.
Switch the lamp on or off by turning the <Control ball>.
Press the <Measurement key>; the display shows Lamp <off> in reverse video.
Select the menu line Return by turning the <Control ball> and press the <Measurement key>
several times to return to the main menu.
6.9.1.2 Channel sensitivities and dark pulse number
Procedure:
Select the menu line Sample by turning the <Control ball>.
Position the measuring aperture on the white field of the calibration card and initiate a measurement
by pressing the <Measurement key>.
The display shows the channel sensitivity values of the individual colors and the dark pulse number
as well as the number of measurements performed up to this time.
Select the menu line Return by turning the <Control ball> and press the <Measurement key>
several times to return to the main menu.
The Channel sensitivities and Dark pulse number should be in the range: